Section 6
Troubleshooting

6.1 Troubleshooting Table

Problem

Cause

Solution

No display with power on

Burned out fuse

Replace fuse.

No RFU values with samples

Lamp OFF

Insure that the lamp is ON.

Lamp burned out

Replace lamp.

Thermal fuse tripped

Call Bio-Rad Service.

Lamp OFF with power ON

Lamp OFF

Insure that the lamp is ON.

Thermal fuse tripped

Call Bio-Rad Service.

Negative values

Readings below set

1. Zero the instrument with the

zero (blank)

range set to 00000. Set the range
with the highest concentration
standard.

2. Zero between gain settings.

Readings fluctuate

Interference in light

1. Check that there are no scratches

path

on the cuvette.

2. Check that there are no particu-

lates floating in the cuvette.

3. Wipe the cuvette walls with a lint-

free tissue.

4. The instrument reading may

fluctuate somewhat at high gain
setting with low concentration
samples.

5. Insure that the gain and range

have been set properly (Sections
4.3 and 4.4).

6.2 Performance Check—Internal Diagnostics

Under normal conditions, the VersaFluor Fluorometer will operate for many years with

minimal maintenance. If the instrument is operating, but a problem is suspected, run the per-
formance check.

1. The VersaFluor Fluorometer instrument can be tested by internal diagnostics which ver-

ify correct operation of the electronics. With the sample compartment door closed, the
internal diagnostics are accessed by holding down any of the four buttons while turning
on the instrument. The diagnostic display below will be shown.

18

DIAGNOSTICS

*****

MED

.06FS

LAMP