Step 5: Reading the ProteinChip PG20 Array in a ProteinChip

SELDI Reader

After the EAM has dried on the array, read the array in a ProteinChip SELDI reader.
The spot protocol used to read the array can be created once and saved. The data
collection parameters of the spot protocol are a function of molecular mass and
need to be adjusted for each antigen. See the operation manual for ProteinChip
data manager software for detailed instructions on how to create an optimized spot
protocol. When optimizing the protocol for the detection of TNF-

α antigen

(molecular mass 17.5 kD), set the Mass Range limits from 0 to 200,000 Da, and
the Focus Mass to 17,500 Da.