Missing reagent notice – Dako Autostainer User Manual

Page 40

Advertising
background image

5. Check that the reagent vials have no bubbles on top of the solution. Remove any

bubbles.

Missing Reagent Notice


The Autostainer displays a prompt when an insufficient volume of reagent is detected.
You can pause the system to add more reagent to the specified vial. If sufficient volume
of a reagent is not available, and you do not respond to the alarm, the Autostainer
continues the run and provides a list of skipped or partially applied slides on the Run
Log.

To program the system to check for sufficient volumes at the start of the run, select the
Check Volumes option on the Set Start Time screen (see Starting a Run, Section 9).

32

Autostainer Handbook

Advertising