CALIBRATION AND SETUP, CONT.

2. CNVT UNITS=X XX (CONVERTED UNITS)

With SCALE 1 SETUP MENU #1 displayed the current setting for the

CNVT UNITS

parameter will be shown. Note that X is the current value and XX is the description
the setting displayed is acceptable, proceed to the next setup parameter.

=

. If

Otherwise, press the 2 key, the ENTER key and then using the numeric keys, enter the
value for the converted units and then press the ENTER key to save the new setting

SCALE 1 SETUP MENU #1. Allowable values are 0 through 6.

U

BASE and CONVERTED UNITS SELECTIONS

and return to

U

0 = none

3 = lb (pounds)

6 = t (tonnes “metric tons”)

1 = tn (tons)

4 =

oz (ounces)

2 =

5 = kg (kilograms

g (grams)

3. Z TR

ACKING RANGE)

With S

displayed the current setting for the

Z TRACK=

parameter will be shown. Note that X.X is the current value. If the setting displayed is
accep

Otherwise, press the 3 key, the ENTER key and then using the numeric keys, enter the

cking range and then press the ENTER key to save the new setting

4.0, 5.0, 6.0, 7.0, 8.0 and 9.0.

4. Z LIMIT=XXX (ZER

With S

r the

Z LIMIT=

param

If the setting displayed is

accep

Otherw

ENTER key, YES or NO (on the soft keys) and then the

ENTE

Z LIMIT=YES

Z LIMIT=NO

Full capacity (no limit)

5.

the

PUO = YES

PUO = NO

Automatic Re-Zero on Power-Up

No Re-Zero on Power-Up

ACK=X.X (ZERO TR
CALE 1 SETUP MENU #1

table, proceed to the next setup parameter.

value for the zero tra
and return to SCALE 1 SETUP MENU #1. Allowable values are: 0.0 (disables Zero
Tracking), 0.5, 1.0, 2.0, 3.0,

O LIMIT RANGE)

CALE 1 SETUP MENU #1 displayed the current setting fo

eter will be shown. Note that XXX is the current value.

table, proceed to the next setup parameter.

ise, press the 4 key, the

R key to save the new setting and return to the

5% of scale capacity

PWR UP Z=XXX (POWER-UP ZERO FEATURE)

With SCALE 1 SETUP MENU #1 displayed the current setting for the

PWR UP Z=

arameter will be shown. Note that XXX is the current value. If the setting displayed is

p
acceptable, proceed to the next setup parameter.

Otherwise, press the 4 key, the ENTER key, YES or NO (on the soft keys) and then

NTER key to save the new setting and return to the

E