CALIBRATION AND SETUP, CONT.

Single-Point for Zero Only (Only Zero) Calibration

. Insure the scale is empty and then press the ENTER key.

TING... for a few seconds and then change to

CAL 2=XXXXXXX. This is the second of two calibration steps.

The calibration parameters (

Swt, Sct and Zct) represent the test load (weight) and the

analog to digital readings at zero and test load. These values are established when the

ration settings (without using

test we

rs values.

st

the setting displayed is acceptable, proceed to the next setup

e

calibrated test load Span weight and then press the ENTER key to save

are

Note that the capacity

ble, proceed to

Otherwise, press the 2 key, the ENTER key and then using the numeric keys, enter the
value for the test load Span count and then press the ENTER key to save the new
setting and return to the SCALE 1 CALIBRATION MENU. Allowable values are 1
through 9,999,999. Note that the capacity can not exceed 7 digits or 9,999,999.

CAL 1= (FIRST CALIBRATION WEIGHT)

With the SCALE 1 CALIBRATION MENU displayed, press the 1 and 0 keys and then the
ENTER key. The display will change to show the settings for capacity, interval, decimal
point position and

CAL 1=0.0. This is the first of two calibration steps.

1

2. The display will flash

CALIBRA

show

CAL 2=.

CAL 2= (SECOND CALIBRATION WEIGHT)

The display will show

1. Press the ZERO key.

2. The display will change to show the SCALE 1 CALIBRATION MENU.

Calibration Parameters

scale is calibrated. To return the indicator to its present calib

ights) enter the calibration paramete

1. Swt=XXXXXXX (SPAN WEIGHT)

With the SCALE 1 CALIBRATION MENU displayed the current setting for the

Swt=

parameter will be shown. Note that XXXXXXX is the current value for the calibrated te
load Span weight. If
parameter.

Otherwise, press the 1 key, the ENTER key and then using the numeric keys, enter th
value for the
the new setting and return to the SCALE 1 CALIBRATION MENU. Allowable values
1 through 9,999,999.

can not exceed 7 digits or 9,999,999.

2. Sct=XXXXXXX (SPAN COUNT)

With the SCALE 1 CALIBRATION MENU displayed the current setting for the

Sct=

parameter will be shown. Note that XXXXXXX is the current value for the analog to
digital reading at test load Span count. If the setting displayed is accepta
the next setup parameter.

8200-M538-O1

y 225 Installation