4 rs-232 data commands for sbi protocol, Rs-232 data commands for sbi protocol – Salter Brecknell SBI100 User Manual

Page 22

Advertising
background image

20

SBI 100 User Instructions Manual

4.4 RS-232 Data Commands for SBI Protocol

The RS232 can be set so a bidirectional connection can be established between the
indicator and the host. To establish this connection, set parameter P4 to 7, and
configure setting P5 (baud) and P6 (parity) to host device. Commands can then be sent
from the host to the indicator using the following commands (ensure the letters entered
are in CAPS) (<CR> - Enter)

Key Symbols

<LF>

Line feed

<CR>

Carriage Return

<ETX>

End of text character

<SP>

Space

H1H2H3

3 status bytes

<p>

Polarity character including minus sign for negative weigh
and a space character for positive.

W1-W7

Weight data

<dp>

Decimal point

U1U2:

Unit measure, kg, lb or oz

Command

Action

Response

W<CR>

Takes a reading

over capacity

under capacity

zero point error

reading (kg or lb)

<LF>^^^^^^^^U1U2 U3U4U5<CR><LF> H1H2H3<CR><ETX>
<LF>________U1U2<CR><LF> H1H2H3<CR><ETX>
<LF>-------- U1U2<CR><LF> H1H2H3<CR><ETX>
<LF><p>W1W2W3W4W5W6<dp>W7U1U2<CR><LF>H1H2H3<CR><ETX>

S<CR>

prints status bytes

<LF> H1H2H3<CR><ETX>

Z<CR>

zeros the scale

<LF> H1H2H3<CR><ETX>

T<CR>

sets up a tare

<LF> H1H2H3<CR><ETX>

U<CR>

changes the units

<LF> U1U2<CR><LF> H1H2H3<CR><ETX>

L<CR>

activates the hold function <LF> H1H2H3<CR><ETX>

X<CR>

switches off the scale

Indicator switches off

?

unrecognised command

<LF>? <CR><ETX>