23

5 Scale Calibration

The scale is configured from the factory to match the specified settings for each unit,
as defined by the product specifications and sales brochure. Modification of the
settings can be accomplished by altering user configuration settings P7-P10.

5.1 Setup Controls

5.2 User Configuration Settings

CAUTION!
Calibration and/or configuration of calibration settings of your scale should be
accomplished by a trained service technician using certified weights to ensure
proper operation and accuracy. Calibration is not covered under warranty.

Moves flashing digit

Change flashed digits

Saves data and move down to next

parameter setting

Exits setup mode

Hold

Print

Units

Tare

On/Off

Zero

kg

kg

Config Menu Avail. Settings

Definition

Detailed Setting

P7

00-31

Displayed Resolution

(00) = 500

(01)= 600
(02)= 750
(03)= 800

(04) =1000

(05)=1200

(06) =1500

(07) = 2000

(08) = 2400
(09) = 2500

(10)= 3000

(11) = 3500

(12)= 4000

(13) = 5000
(14) = 6000

(15) = 7000
(16) = 7500
(17) = 8000

(18) = 10,000

(19) = 12,000
(20) = 15,000

(21-31) = N/A

P8

0,1,2

Division Increment

0 = 1; 1 = 2; 2 = 5

P9

0-5

Decimal Position

0 = 123456; 1 = 12345.6; 2 = 1234.56;
3 = 123.456; 4 = 12.3456; 5 = 1.23456

P10

0,1

Calibration weight

0 = kg; 1 = lb