7 error codes – Salter Brecknell SBI100 User Manual

Page 31

Advertising
background image

SBI 100 User Instructions Manual

29

7 Error Codes

Error Message

Definition

Required Solution/Troubleshooting

0¯ ¯ ¯ ¯:

Weight above range for calibrated
zero point.

Remove load before zeroing Or Recalibrate the scale.

0_____:

Weight below range for calibrated
zero point.

Remove load before zeroing Or Recalibrate the scale.

_____:

Indicates an under range condition

Remove all loads and zero the scale.

¯ ¯ ¯ ¯:

Capacity exceeded

Remove the load and try again. If the load is great than 6 kg a scale
with a larger capacity is required.

CAL-Er:

Calibration error

Restart calibration

Lo.bAt:

Low Battery

Recharge the battery. Upon initial use, it is recommended to charge
battery for 8 hours prior to use.

EEP.E0

EEPROM can't be accessed

Replace S100 Indicator

EEP.E1

Configuration settings have changed
and not been stored

Calibrate the scale to store settings

EEP.E2

P7-P9 settings exceed scale's normal
range

Reconfigure P7-P9 setting