SBI 100 User Instructions Manual

5.3 Calibrate Scale

You may choose to configure your scale for a higher resolution. The factory does not
recommend increasing the resolution above 7,500 divisions for a stable weight reading.
Certain environments may cause the scale to be unstable at factory settings, reduce
the # of division settings to increase your stability.

Calibration can be done with 25% to 100% of requested load and can be calibrated with
1 or 2 calibration points

1.

2.

3.

4.

Enter in calibration weight from 25% to 100% of full capacity

5.

Single point calibration, enter the same weight in again and move to step 7. For 2 point calibration enter in the second
calibration weight between 25% 100% full capacity.

6.

On/Off

Zero

Tare

Tare

Tare

25%-100%

kg/lb

50.00

50.00

Hold

Print

Units

Tare

Tare

25%-100%

kg/lb

50.00

100.00

Hold

Print

Units

Tare