SBI 100 User Instructions Manual
5.3 Calibrate Scale
You may choose to configure your scale for a higher resolution. The factory does not
recommend increasing the resolution above 7,500 divisions for a stable weight reading.
Certain environments may cause the scale to be unstable at factory settings, reduce
the # of division settings to increase your stability.
Calibration can be done with 25% to 100% of requested load and can be calibrated with
1 or 2 calibration points
1.
2.
3.
4.
Enter in calibration weight from 25% to 100% of full capacity
5.
Single point calibration, enter the same weight in again and move to step 7. For 2 point calibration enter in the second
calibration weight between 25% 100% full capacity.
6.
On/Off
Zero
Tare
Tare
Tare
25%-100%
kg/lb
50.00
50.00
Hold
Units
Tare
Tare
25%-100%
kg/lb
50.00
100.00
Hold
Units
Tare