Salter Brecknell HS-300 User Manual

Page 11

Advertising
background image

16

x = 5 : Power-on zero-point ±10%FS;
x = 6 : Power-on zero-point ±20%FS;
x = 7 : No limitation
x = 8 : Power-on zero-point +1%FS
x = 9 : Power-on zero-point +2%FS
x = 10 : Power-on zero-point +3%FS
x = 11 : Power-on zero-point +4%FS
x = 12 : Power-on zero-point +5%FS
x = 13 : Power-on zero-point +10%FS
x = 14 : Power-on zero-point +20%FS
If zero key can be activated, it can clear the tare weight. If zero point is above
the setting range, the indicator will show “0¯ ¯ ¯ ¯”, and if zero point is below
the setting range, “0_ _ _ _” will be shown.

5.14) P14.x: select which zero point will be used when powered on and weight

signal is within the power-on zero-point range:
x = 0 : Choose current weight as current power-on zero point.
x = 1 : Choose calibration zero point as power-on zero point
x = 2 : Choose power-off zero-point as power-on zero point, and power-off

tare weight as current tare weight.

5.15) P15.x: select which zero point will be used when powered on and weight

signal is NOT within the power-on zero-point range:
x = 0 : Choose current weight as current power-on zero point.
x = 1 : Choose calibration zero point as power-on zero point
x = 3 : Choose power-off zero-point as power-on zero point, and power-off

tare weight as current tare weight.

5.16) P16.x: Zero tracking range

x = 0 : 0d, means no tracking
x = 1 : +0.25d
x = 2 : +0.5d
x = 3 : +1d
x = 4 : +1.5d
x = 5 : +2d
x = 6 : +3d
x = 7 : +4d

17

x = 8 : +5d
Choose the zero tracking range as per the stability of weighing system,
accuracy and weight excursion. The normal setting is +0.5d ~ +1.5d.

5.17) P17.x: Data filter intensity

x = 0 very weak
x = 1 : weak
x = 2 : middle
x = 3 : strong
The larger the chosen digit is, data filter intensity is stronger, and the speed
of data updating is lower. The normal setting is 2.

5.18) P18.x: Range of weight stability checking

x = 0 : +0.5d
x = 1…9 : +1d --- +9d
If the variety of weight reading is within the setting range continuously for
several times, the scale is recognized as stable. The normal setting is +1d ~
+3d.

5.19) P19.x: Overload range that can be displayed (when weight is lager than

range, “ˉˉˉˉˉ”will be shown):
x = 0 : FS+0d
x = 1 : FS+9d
x = 2 : 101%FS
x = 3 : 102%FS
x = 4 : 105%FS
x = 5 : 110%FS
x = 6 : 120%FS
x = 7 : 150%FS
x = 8 : 200%FS
x = 9 : No limitation

5.20) P20.x: Backlight on-off mode can be displayed

x = 0 : Backlight is always off
x = 1 : Backlight is always on
x = 2 : Backlight is auto on and auto off. It is auto off after 10s when scale
goes to stable and has no key operation, and it is auto on when scale is
unstable or there’s some key operation.