·OPERATION INTERFACE:
1. Display: 0.65” (17mm) 7-segment LCD, 5

1

/

2

digits

2. Keyboard: 4-key push button

·POWER:
1. Alkaline Batteries: 4 x “AAA” size cells

When all displayed segments of LCD flashed, this indicates the batteries are
low below 4.9V and you’d better to replace batteries;
When “Lo.bAt” displayed, this indicates the batteries are low below 4.7V and
you should replace batteries immediately.

2. AC Adapter: 6VDC, 500mA, with central negative:
3. Work current: ≤25mA

(when voltage in 5Vdc-8Vdc and not include load cell’s consumption)

·OPERATION TEMPERATURE: 20℃±15℃
STORE TEMPERATURE: -10℃-70℃
OPERATION HUMIDITY: ≤95%RH (no condensate)

·LOADCELL:

Because of more than one load cell can be used on a scale, following are

required on the load cell set to be used with this indicator,
1. Sensitivity: 0.3mV/V --- 3mV/V (must be fit to >0.2uV/display grad)
2. Input Resistor: ≥80 Ω
3. Output Resistor: <10 KΩ

·LOADCELL WIRING:
PIN 1:
RED, EXCITATION +

PIN2: BLACK, EXCITATION –

PIN3: GREEN, SIGNAL -

PIN4: WHITE, SIGNAL +

3

.

KEY FUNCTION:

1.

FACEPLATE:

2.

DISPLAY SYMBOL MEANING:

Zero ◄--------The scale is at zero point and the gross weight is 0

Net ◄---------The display reading is net weight, and the tare weight is not 0.

Hold -------The scale is under HOLD mode.
It displays the current live weight when flashed, and the locked reading will
be shown when  does not flash and comes steady.

3.

SUMMARY of KEY DEFINITION:

3.1 In normal weighing mode press and hold down keys for 3 seconds:




(1)


a. If this key is only set for HOLD (P2=0), press this key to enter or exit HOLD

mode.

b. If this key is only set for PRINT (P2=1), press this key to output the data

according to P4 setting.

c. If this key is set for both HOLD and PRINT function (P2=2), press this key to

output the data according to P4 setting.

HOLD

PRINT

UNIT

TARE

ZERO

HOLD

PRINT