Probe specifications – ETS-Lindgren HI-3804 RF Industrial Compliance Meter User Manual

Page 12

Advertising
background image

12

Specifications

www.ets-lindgren.com

Probe Specifications

Frequency Range:

10 MHz

–42 MHz

Frequency Bands:

Three:

10 MHz

–20 MHz

20 MHz

–35 MHz

35 MHz

–42 MHz

Measurement Range:

Three ranges:

0.5 mW/cm

2

–50.0 mW/cm

2

Probe Overload:

1 mW/cm

2

Calibration Frequencies:

13.56 MHz

27.12 MHz

40.68 MHz

Dimensions:

Length: 35.6 cm (14.0 in)

Probe cover diameter:

13.2 cm (5.2 in)

Weight:

0.28 kg (0.62 lb)

Environmental

Operating Temperature:

10

°

C to 40

°

C

50

°

F to 104

°

F

Humidity:

0% to 90% relative humidity,

non-condensing