BAT

Displays battery status. After BAT is pressed, the battery voltage displays for

two seconds and then the readout returns to measurement mode. When the

battery voltage drops below a preset limit, the battery symbol will blink, indicating

that the batteries need replacing. If the battery voltage is allowed to drop below

the limit required for proper device operation, the LCD will turn blank.

UNITS

Instructs the probe to change the units of measure and displays the new units.

When measuring the electric field, UNITS toggles between mW/cm

2

and V/m. In

the magnetic field, it toggles between mW/cm

2

and mA/m.

CLEAR

DATA

Clears all readings from datalog memory. To perform this function, press and

hold CLEAR DATA. The characters clr will flash on the display for two seconds.

Continue pressing CLEAR DATA until 000 appears, indicating that datalog

memory is cleared. Release the key to return the readout to measurement mode.

Note: If CLEAR DATA is released while the clr characters are

flashing, datalog memory is unaffected. This helps prevent

accidental erasing of data.

LOG

Saves the current measurement in datalog memory. As this occurs, the

three-digit identification number of the reading momentarily displays. The data

stored includes values, units of measure, over/under range indication, and active

axes. Up to 150 measurements may be stored. When datalog memory is full,

pressing LOG replaces the value stored in location 150 with the new value; all

other memory locations are unchanged.

Note: If LOG is pressed for longer than two seconds, the data just

logged is displayed.