Mode select, Range – ETS-Lindgren HI-3804 RF Industrial Compliance Meter User Manual

Page 16

Advertising
background image

16

Operation

www.ets-lindgren.com

MODE

SELECT

The keypad uses a matrix for the top three rows of keys. The function of each

key depends on the location of the cursor in the LCD. When the readout is turned

on, the cursor is located in the first of the four function columns. To access a

function in this mode, press the corresponding arrow key. For example, to toggle

between electric and magnetic field measurements, press the lowest arrow key.

Press MODE SELECT to move the cursor to the right, advancing to the next

column. Each time MODE SELECT is pressed, the cursor moves another

position to the right, and then returns to the first position. This matrix allows a

total of 12 functions to be assigned to the top three rows of keys. Following is a

description of each of the functions.

RANGE

This changes the fixed

ranges or scales. When

turned on, the HI-3804

is in AUTO RANGE

mode, and determines

the correct range within

the current mode (E or

H field) according to the

detected field level. As

the field being

measured increases or

decreases, the range is

automatically selected

for best resolution and

accuracy.

In some situations it may be helpful to fix the scale of the HI-3804. Press RANGE

once to select the current scale setting. Each time you press RANGE the scale

moves to the next least sensitive range. When the least sensitive scale is

reached, it starts again at the most sensitive range.