Prev, Next, Freq – ETS-Lindgren HI-3804 RF Industrial Compliance Meter User Manual
Page 19
www.ets-lindgren.com
Operation
19
PREV
Accesses the last value stored in datalog memory. When PREV is pressed, the
three-digit identification number of the stored value displays for one second and
then the stored value displays. This value displays as long as the PREV key
remains pressed. Two seconds after releasing the key the readout returns to
measurement mode. Pressing the PREV key at that point decrements the
displayed value toward the beginning of memory (value 001). If PREV is pressed
while viewing value 001, the readout wraps around to the highest stored
identification number.
NEXT
Accesses the next value stored in datalog memory. NEXT is similar to the PREV
key, except that when you continue to press NEXT, it increments the displayed
value toward the end of datalog memory. Additionally, if NEXT is pressed while
viewing the highest stored value, the readout wraps around to identification
number 001.
Note: If an under range value is logged and displayed, the
measured value and the under range indicator will appear. If an
over range value is logged and displayed, an over range alarm will
sound and the over range indicator (
►
) will appear.
FREQ
Allows the frequency to be changed by the user. When pressed, the display
reads F-1 (13.56 MHZ), F-2 (27.12 MHZ), or F-3 (40.68 MHZ). These values can
also be found in the lower right corner on the front of the readout.