ETS-Lindgren HI-3804 RF Industrial Compliance Meter User Manual
Page 17
www.ets-lindgren.com
Operation
17
To return to AUTO RANGE mode, press and hold RANGE for two seconds until
AUTO displays.
MAX
Note: The maximum value is not affected by the CLEAR DATA key.
The maximum value is cleared when:
The readout is powered up.
The MAX key is released.
The unit of measure is changed.
MAX displays the maximum reading. During field measurements, the processor
continually monitors and stores the highest measured field value. To recall and
display this value, press MAX. The value displays as long as MAX is pressed.
When MAX is released, the value displays for two seconds, and then the MAX
memory location is cleared and a new maximum reading is accumulated.
Note: If the maximum reading is over full scale for the range, an
over range alarm will sound and the over range indicator (
►
) will
display.
E/H
Toggles the measurement mode between electric and magnetic field readings.
The current measurement units are displayed to the right of the reading. For
E field measurements, V/m will illuminate; for H field measurements, A/m will
illuminate.
Note: The field being measured when the HI-3804 was last turned
off becomes the default field when the HI-3804 is turned on again.