Introduction

The HI-6113 is an automated laser controller that accepts commands from the

software running on the computer and then returns the requested data. Upon

receiving the command to turn on the system, the HI-6113 initiates a startup

sequence that turns on the laser, which provides power to the Field Probe. The

Field Probe, when powered up and running, continuously sends data to the

HI-6113. The HI-6113 returns data to the computer software upon request in

either unprocessed raw form or processed form, depending on the requested

data format. The USB interface provides the data communication between the

computer software and the HI-6113.

The Field Probe system incorporates a safety interlock mechanism that turns off

the laser if the HI-6113 does not receive data from the probe within a specified

time frame. The safety mechanism is intended to prevent injury from the laser if

the HI-6113 issues a command to turn on the laser while the fiber optic cables

are disconnected, improperly connected, cut, or damaged.

The universal input power supply provides 5 volts DC power to the HI-6113.

Minimum Computer Requirements

To install ProbeView Laser and operate the Field Probe and HI-6113, you must

have an Intel® Pentium® III computer with one USB port, installed with one of

the Microsoft® Windows® operating systems; see page 18 for a list of supported

operating systems. An externally powered USB hub may be required for

computers without a USB port.

Due to the graphics in ProbeView Laser, a 300 MHz processor or

better is desirable. A slower machine will run ProbeView Laser, but

the sample rates will be slower. Toggling off the bar and scatter

graphs may increase sample rate performance.