Page — 14
HI
n
3550 Manual
The HI
n
3550 can display integrated magnetic flux levels
of up to 999.99 T-h. Integrated flux levels in excess of
this value are not displayed.
5.1 Using the Integrating (
II
B) Mode
1. Slide the AC-DC (Magnetic Field Selection) switch to
select the type of magnetic field to be measured.
Set the STBY-
I
B-
)
B (Measurement Mode) Switch to
I
B.
NOTE
Do not perform the next step if you wish to add
to previously stored data.
2. Press the SELECT and SET buttons (CLEAR function)
on the back of the instrument simultaneously for
approximately three seconds to clear the memory of
stored integrated time and value.
The alarm settings remain unchanged after the memory
is cleared.
Changing the Measurement Mode
Changing from integrating to instantaneous measurement
mode will stop the integrating measurement mode while
retaining the integrated time and value. After you change
the measurement mode, the reading will remain
unchanged until one measurement cycle (3 seconds) is
completed.