Page — 16
HI
n
3550 Manual
Figure 7
Stand-by Mode
value that can be displayed is 3.54 T. These readings are
also displayed for any measured field strength exceeding
these values.
5.3 Changing the Target Magnetic Fields
To change the target magnetic field while measuring,
slide the AC-DC switch to the desired position.
NOTE
Be very careful to select the appropriate setting
(AC or DC) for the magnetic field being
measured. Incorrect settings can increase the
error or uncertainty of the displayed value by up
to 10%.
5.4 Using the Stand-by (STBY) Mode
To stop measuring, change the measurement mode to
STBY. Selecting STBY reduces power consumption and
prolongs battery life. All stored integrated field data is
saved in STBY mode. For practical purposes, the STBY
mode is the "OFF" condition of the instrument. If the
instrument will not be used for a considerable period,
remove the batteries.