Cleaning disinfection, and sterilization, Cleaning – Welch Allyn KleenSpec Vaginal Specula Illumination System - User Manual User Manual

Page 10

Advertising
background image

6

Cleaning Disinfection, and Sterilization

Cleaning

Wipe

Transformer (REF 73405 or REF 73410), Illuminator (REF 78800), Expendable
Illuminator (REF 78600)

Refer to Cleaning Table to identify the appropriate cleaning product.
1.

Disconnect the transformer from the wall outlet and then disconnect the
illuminator from the transformer.

2.

Disconnect illuminator from expendable illuminator and remove the

expendable illuminator from the speculum (58600/58601 only).

3.

Follow the disinfecting wipe manufacturer’s instruction and wipe cord,
illuminator, or expendable illuminator completely to remove any residue.

4.

This process is a rough clean only. Refer to disinfection or sterilization for
further processing.

Soak

Refer to Cleaning Table to identify the appropriate cleaning product.

Illuminator (REF 78800) and/or Expendable Illuminator (REF 78600)
1.

Disconnect the transformer from the wall outlet and then disconnect the
illuminator from the transformer.

Illuminator

REF 78800

Transformer

REF 73405 or

REF 73410

Expendable

Illuminator

REF 78600

Cleaning

CaviWipe
(Wipe)

Yes

Yes

Yes

Enzol (Soak)

Yes

Limited*

Yes

Metrizyme
(Soak)

Yes

Limited*

Yes

Disinfection

Caviwipe

Intermediate

Level

Limited*

Intermediate

Level

Yes

Intermediate

Level

Sterilization

Autoclave

No

No

Yes**

* Do not immerse the transformer plug. Follow instructions per this manual.

** The Expendable Illuminator’s ability to transmit light will degrade after

repeated autoclave cycles (approximately 20) and will eventually need
replacement.

Lamp must remain installed during all cleaning and disinfection
processes.