8

3.

Thoroughly rinse the cord only with 22° C to 43° C tap water to remove
disinfectant.

4.

Allow to dry.

Sterilization

Refer to the Cleaning Table to identify the appropriate sterilization method/prod-
uct for the expendable illuminator.

Wet Steam Autoclave

Expendable Illuminator (REF 78600)
1.

Clean expendable illuminator as instructed in “Cleaning-Soak”.

2.

Place the expendable illuminator in the autoclave chamber to allow best
exposure to steam. Autoclave chamber parameters are as follows:

Type:

132° Pre-vacuum (Hi-VAC) Steam Cycle

Exposure Temperature:

132°C to 135°C (270°F to 275°F)

Exposure Time:

3 - 5 minutes

Exhaust Time:

Fast

Disposal

There are no harmful materials associated with the Illumination System.

Device is considered a biohazard after use. Follow good hospital practices and
local ordinances in the disposal of the illuminator.

There are no risks associated with the disposal of this device after its useful life if
it is cleaned and/or disinfected per the instructions in this manual.