Disinfection – Welch Allyn KleenSpec Vaginal Specula Illumination System - User Manual User Manual

Page 11

Advertising
background image

7

2.

Disconnect illuminator from expendable illuminator and remove the

expendable illuminator from the speculum (58600 / 58601 only).

3.

Follow the cleaning solution manufacturer’s instructions and soak the
illuminator and/or expendable illuminator as instructed. Solution temperature
is 22° C to 43° C.

4.

Scrub any area of visible debris with a soft brush and thoroughly rinse the
illuminator and/or expendable illuminator with 22° C to 43° C tap water to
remove the cleaning solution.

5.

Rough dry the illuminator and/or expendable illuminator.

6.

The illuminator and/or the expendable illuminator is now ready for
disinfection.

Limited Soak

Transformer (REF 73405 or REF 73410)

Refer to Cleaning Table to identify the appropriate cleaning product.
1.

Disconnect the transformer from the wall outlet and then disconnect the
illuminator from the transformer.

2.

Follow the cleaning solution manufacturer’s instructions and soak the cord
only
in the solution as instructed. Solution temperature is 22° C to 43° C.

3.

Scrub any area of visible debris with a soft brush and thoroughly rinse
transformer cord only
with 22° C to 43° C tap water to remove the cleaning
solution.

4.

Rough dry transformer.

5.

The transformer is now ready for disinfection.

Disinfection

Refer to the Cleaning Table to identify the appropriate disinfection method/prod-
uct for the illuminator, expendable illuminator, and/or transformer.

Illuminator (REF 78800) and/or Expendable Illuminator (REF 78600)
1.

Clean illuminator and/or expendable illuminator as instructed in “Cleaning-
Soak” or “Cleaning-Wipe”.

2.

Follow the disinfecting wipe manufacturer’s disinfection instructions.

3.

Thoroughly rinse illuminator and/or expendable illuminator with 22° C to 43°
C tap water to remove disinfectant.

4.

Allow to dry.

Transformer (REF 73405 or REF 73410)

DO NOT IMMERSE THE TRANSFORMER PLUG.
1.

Clean transformer as instructed under “Cleaning-Limited Soak” or “Cleaning-
Wipe”.

2.

Follow the disinfecting wipe manufacturer’s disinfection instructions on the
cord only.

Lamp must remain installed during all cleaning and disinfection