Error indications and interpretation, Error codes, 8 - error indications and interpretation – Welch Allyn Spot Vital Signs - User Manual User Manual

Page 45

Advertising
background image

8

37

Error Indications and Interpretation

The following table of conditions and error codes provides a quick reference of the
descriptions and probable causes of error codes.

To clear the error code:

Power the Spot Vital Signs off, wait five seconds, and power on. If the error code
reappears then power the Spot Vital Signs off and disconnect the battery for five minutes.
Reconnect the battery and power on. If the error code continues to reappear, call Welch
Allyn for an RMA Number (see “Technical Assistance” on page 56).

Press the Blood Pressure Start/Stop button to reset flashing patient alarm conditions.

Error Codes

Table 1. General

Code

Description

Corrective Action

E11

Internal safety violation

Check patient, contact Technical Service.

C12

Ambient temperature out of range

Adjust ambient temperature or device location.

C13

Battery failure

Use wall transformer.

E0.0 - E9.9

Temperature module malfunction

Contact Technical Service.

E42

Internal communications error

Disconnect the battery and wait 5 minutes.
Reconnect the battery and then set the date and time,
see “Blood Pressure Hose and Cuff Connections” on
page 21.

E20 - E50

General internal malfunction

Contact Technical Service.

Table 2. Blood Pressure

Code

Description

Corrective Action

C02

Auto-zero failure

Check for air obstruction, limit patient movement.

C03

Inflation too rapid

Check for kinked blood pressure cuff tubing, pressure
hose, or other air obstruction.

C04

Excessive inflation time

Check for air leaks.

C05

Excessive noise

Check patient condition, blood pressure cuff
placement, limit patient movement.

C06

Measurement was outside of device’s
measurement range

Check patient condition.

E10

Blood pressure cuff overpressure condition

Check patient condition.