Slide the new battery completely into the compartment. Lay the connector on the
battery. The relief pocket in the battery door purposely provides sufficient clearance for
the battery connector.

1.

Replace the battery door and tighten each of the 4 screws.

2.

Connect the AC power transformer to the Spot Vital Signs and charge the new battery
for 16 hours. It is possible to use the Spot Vital Signs during this charging period.

The battery is a lead-acid battery. In the USA, call 1-800-SAV-LEAD for instructions on how
to recycle. For users outside the US please contact your local authorities on recycling.

Masimo SpO

2

Calibration Check

Use a Masimo-approved SpO

2

simulator (Fluke Biotek Index 2 or Clinical Dynamics

SmartSat) to check the SpO

2

accuracy. There is no way to change the calibration of the

SpO

2

module. If the SpO

2

is out of calibration, contact Technical Service.

Nellcor SpO

2

Functional Check

Use a Nellcor SpO

2

simulator (SRC-MAX ) to check the SpO

2

functionality. There is no

way to change the functionality of the SpO

2

module. If the SpO

2

is not functioning

properly, contact Technical Service.

SpO

2

Accessory Disposal

Dispose of all finger sensors and cables in accordance with facility, local, and goverment
regulations.

Temperature Calibration Check

Use the 9600 Plus Calibration Tester to check the SureTemp thermometer accuracy. If the
thermometer is out of calibration, contact Technical Service.

Relief Pocket

Battery Connector

Battery