Mechanical, Electrical, Environmental – Welch Allyn Spot Vital Signs - User Manual User Manual

Page 54: Mechanical electrical environmental

Advertising
background image

46

Welch Allyn Spot Vital Signs

Mechanical

Electrical

Environmental

Dimensions

Height: 9.7 inches (24.6 cm)

Length: 5.7 inches (14.5 cm)

Depth: 4.7 inches (12.0 cm)

Weight

Approximately 4.6 pounds (2.2 kg)

Mounting

Self-supporting on rubber feet

Custom Mobile Stand

Custom Wall Mount

Custom IV Pole Mount

Portability

May be hand-carried when held by the rear handle.

Power Requirements

Patient-rated isolation transformer is connected to AC mains:

North American Version: 120VAC, 60Hz. 0.13A Input, 7.2VDC, 1.0A Output

International Version: 230VAC, 60HZ, 0.065A Input, 7.1VDC, 0.860A Output

Australian Version: 240VAC, 50Hz, 13VA Input, 7.2VDC, 1.0A Output

Battery

Lead acid, with external charger.

A fully charged battery supports 130 typical blood pressure determinations taken at

7-minute intervals. The battery is 90-100% charged after 12 hours of charging. The

battery automatically charges when the Spot Vital Signs is powered through the AC

power transformer. The battery charges faster when the instrument is not in

operation.

Operating Temperature

+10

 to +40C (Thermometer operating temperature 16° to 40° C)

+50

 to +104F (Thermometer operating temperature 61° to 104° F)

Storage Temperature

-20

 to +50C

-4

 to +122F

Transport Temperature

-20

 to +49C

-4

 to +122F

Relative Humidity

15 to 90% (non-condensing)

Operating Altitude

-170 to +4877 m

-557 to +16,000 ft.