Nellcor® sensor accuracy guide, Nellcor patents, Nellcor – Welch Allyn Spot Vital Signs - User Manual User Manual

Page 53: Sensor accuracy guide

Advertising
background image

Directions for Use

45

Nellcor

®

Sensor Accuracy Guide

Accuracy specifications are based on controlled hypoxia studies with healthy,
non-smoking adult volunteers over the specified saturation SpO

2

range. Pulse oximeter

SpO

2

readings were compared to SaO

2

values of drawn blood samples measured by

hemoximetry. All accuracies are expressed as + “X” digits. This variation equals plus one
standard deviation (+ 1 SD), which encompasses 68% of the population.

Nellcor Patents

Covered by one or more of the following U.S. patents and foreign equivalents:

5,485,847; 5,676,141; 5,743,263; 6,035,223; 6,226,539; 6,411,833; 6,463,310; 6,591,123;
6,708,049; 7,016,715; 7,039,538; 7,120,479; 7,120,480; 7,142,142; 7,162,288; 7,190,985;
7,194,293; 7,209,774; 7,212,847; 7,400,919

.

Pulse Rate Range

25 to 240 bpm

Pulse Rate Accuracy

±3 bpm

±3 bpm (low perfusion)

Table 15. OxiMax Sensor Models, Single Patient Use

Sensor Models

SpO

2

Range

70% to 100%

MAX-AI

± 2

MAX-PI

± 2

MAX-II

± 2

MAX-RI

1

± 3.5

1

The accuracy specification has been determined between saturations of 80% to 100%.

Table 16. OxiCliq Sensor Models, Single Patient Use

Sensor Models

SpO

2

Range

70% to 100%

OXICLIQ-PI

± 2.5

Table 17. Reusable Sensor Models

Sensor Models

SpO

2

Range

70% to 100%

D-YS (Infant to Adult)

± 3

D-YS and D-YSE

± 3.5

D-YS and D-YSPD

± 3.5

DS-100A

± 3

OXI-A/N (Adult)

Adult: ± 3

OXI-P/I (Pediatric/infant)

± 3