PACKING LIST

SECTION II: Airtec Kit Components

12

NEXTTEQ LLC (Rev. 2) 4/09



1.2 AIRTEC KIT

Your Airtec Kit includes the following components:

Regulator, Regulator Gauge, Flow Meter, Tube Holder


1.3 AIRTEC DETECTOR TUBES

The Airtec Kit is designed for use with the following Gastec Tubes:

Carbon Monoxide, Carbon Dioxide, Nitrogen Oxides, Oil Mist, Water Vapor

WARNING

Use only the Gastec tubes listed above.


Airtec detector tubes are designed to meet the monitoring requirements for maintaining
Grade D (Compressed Gas Association) breathing air. These standards are listed in Table
1 (below).

NOTE

The Airtec Kit is not equipped to monitor for oxygen deficiency. Nextteq LLC
recommends the use of a portable sample draw oxygen meter.

Table 2 (page 13) lists the data for Gastec Airtec Tubes included in the kit. Tube
packages contain detailed instruction sheets on the storage and use of the tubes, as well as
instructions on interpreting tube readings. Tubes are packaged ten per box, allowing ten
tests. Tubes may be used only once.

TABLE 1

COMPRESSED GAS ASSOCIATION GRADE D STANDARDS

Gas/Substance

Level

Comments

Carbon Dioxide

1000 ppm

Maximum level

Carbon Monoxide

10 ppm

Maximum level

Oil Mist

5 mg/m

3

Maximum

level

Oxygen

19.5 – 23.5%

Water Vapor

---

Varies with use requirements


SECTION I: Introduction