Tips for Tube Reading

The following items will be helpful when reading and interpreting the Airtec detector

tubes.

1.) Always read the tubes and record the results immediately after testing. In some

cases stains may fade or “crawl” within a few hours.

2.) A tube may never be reused, even after a zero reading.

3.) The oil mist tube is normally used at 500 scc/min for 40 minutes. However,

reliable results can be obtained using shorter sampling periods and correcting the

reading with a multiplication factor (see below):


Test Duration

Multiply Reading By

20 minutes

2

4.) If at the end of a test you note that the stain has overshot the scale, the actual

concentration may be determined as follows:

a) Repeat the test with a new tube and observe the time required to reach the

highest calibration mark (full scale).

b) Multiply the full-scale concentration by the ratio of the recommended test

time to the actual test time.

Example

H2O tube reaches full scale (80 mg/m

3

) in 36 seconds. Recommended

time is 1 minute.

Ratio = 1 min/0.6 mins. = 1.66

Thus, the actual concentration is

80 x 1.66 = 133 mg/m

3