Calibration procedure – Leica Geosystems Laser Locator User Manual
Page 31
Advertising
![background image](/manuals/651826/31/background.png)
31
Laser Locator-1.0 en
Calibration procedure
Settings
4x
i
Point the Laser Locator roughly
northwards.
Click the azimuth key four times
in rapid succession.
"FIEL Co" appears briefly,
followed by "4 Pt Co" for the
regular 4 point calibration.
Only click the distance key
if you want to perform the
special 12 point calibration.
"12 Pt Co" is displayed.
The selected calibration
procedure begins in a few
moments.
Move the Laser Locator according
to the displayed instructions.
After the last instruction,
analysis begins and the pointing
circle blinks for 4 sec. (4 Pt Co)
or 30–60 sec. (12 Pt Co).
You then see
- "Good Co": indicates
successful calibration.
- "bAd Co" followed by "rES Co":
indicates that the calibration
data are unusable.