BUCHI Syncor Polyvap User Manual

Page 33

Advertising
background image

Syncore

®

Accessories

version E

31

Figure 39: Hose coupling diagram

For sensitive reactions where traces of steam or other ga-
ses can cause disturbances, we recommend the following
procedure to efficiently exchange the atmosphere by an
inert gas:

1. Lock the inert gas inlet a and outlet b on the inert gas

module using a clamp bride or a 3-way faucet.

2. Evacuate the Syncore to 300 mbar.

3. Ventilate the system with inert gas via the Vacuum

Controller (connect the ventilation valve to the inert gas
main).

4. Repeat steps 2 and 3 three times.

5. Open the inert gas inlet a and then outlet b at the inert

gas module.

6. Set the light inert gas flow to c. 1 to 2 bubbles/se-

cond.

Working under inert gas

Legend ------- Electrical connections

Valve or clamping buckles

Vacuum connections
Inert gas connections

Valve Unit

Receiving flask

Rack R-24

Reflux module/

Inert gas module

a

b

Cooling
Unit

Vacuum Controller

Gas washing bottle

Inert gas connection

Inlet

Outlet

3 Function,putting into operation, operation and maintenace

3 Function,putting into operation, operation and maintenace