3.8 Reflux module/flushback module

The reflux module/flushback module a enables simulta-

neous condensation of solvent vapor in glass vessels b (=

reflux*) via contact of the glass vessels with actively cooled
metal ring, which surrounds each glass vessel.
*reflux module R-96, 48, 24 / flushback module R-12, 6,
4

The tubes for the cooling medium must be secured with
tube safeties as illustrated. It is irrelevant what connection
is used for adding or removing cooling media.

The cooling medium flow (tap water or actively cooled water)
must be controlled. The cooling medium flow should not
be greater than 2 liters/min. An excessive pressure on the
silicone tubes must be prevented, because the tubes can
burst and cause material and personal damage.

The reflux module/flushback module cannot be used wi-
thout secured tube connections.

Figure 43: Reflux module on rack R-24 with test tubes

Figure 44: Tubes on reflux module