Warning – BUCHI SpeedDigester K-436 User Manual

Page 29

Advertising
background image

6 Operation

29

K-425 / K-436 Operation Manual, Version A

6 .2 .2

Peroxide digestion —preparing sample tubes

During the preparation process, handling with peroxides and other dangerous substances is inevi-
table. Peroxides are strong oxidants and can form explosives! All steps must be performed under safe
laboratory conditions.

!

WARNING

Death or serious injuries by toxic and explosive peroxides.

Use only damage free sample tubes

•

Exchange worn out or defective parts immediately

•

Handle peroxides in a fume hood

•

Do not bring peroxides into contact with highly combustible material

•

Prerequisites

All sample tubes must be clean and free of damages.

•

To achieve a reliable and reproducible digestion result the quality of sample material and the

•

peroxide must be free of impurities and of adequate amount.

Filling of a sample tube

Take the sample tubes sequentially and weigh in sample material.

•

Carefully(!)

•

add diluted sulfuric acid (e.g. 69%) of sufficient amount to the samples.

Carefully(!)

•

add peroxide (e.g. H

2

O

2

30%) of sufficient amount to the samples.

➡ The sample tubes can now be installed in a rack.
Note

Note down the sample weight and position to keep the probe processing and its analysis repro-

•

ducible. All racks are marked on top with a "1" for the first sample position.
The rack(s) can be used as a temporary storage for empty tubes during the preparation process.

•

When adding the sulfuric acid and peroxide, rotate the tube to wash down any sticky sample

•

material to the bottom and to coat the inside of the tube.

6 .2 .3

Installing a suction module and sample tubes (300 ml)

To withdraw hazardous fumes and vapors a suction module must be installed.

Prerequisites:
All hoses and connectors must be unclogged.

Installation steps:
For installation and handling all parts must be < 40 °C!

Install the plastic adapter piece to the suction module

•

1.

Mount the hose connector at the EPDM hose

•

2 of the suction module (i.e. Scrubber B-414 or

water jet pump).
The connector at the EPDM hose can be stored temporarily at the magnetic bracket at the rear

•

side of the SpeedDigester.

Advertising