6 Operation

32

K-425 / K-436 Operation Manual, Version A

6 .2 .4

Installing 500 ml sample tubes

Unlike 300 ml (and 250 ml) sample tubes, 500 ml tubes can not simply be inserted into the appro-
priate rack. The information in this section serves to set up this type of sample tube properly and
safely in the rack.

Installation steps:

For installation and handling all parts must be < 40 °C!

*CL

ICK

*

1

2

3

frontside

rear side

Plug the rack into the carrier

•

1.

➥ Two audible *click* sounds will appear as soon as the rack is securely locked at the carrier!

➥ Check the locking by lifting the rack together with the carrier (carrier must follow the rack)
Insert the prepared sample tubes into the rack(s)

•

2— start with the first (front) position.

Install the rake

•

3 to interlock the sample tubes at their head side. Check for secure interlocking!

To separate the rack from the carrier, pull the unlock buttons

•

4 on both sides of the rack and lift it

off the carrier.

!

WARNING

Risk of death or serious chemical burns by acids or peroxides when rack tilts.

Do not use faulty parts

•

Install the rack on an even and stable surface

•

Support the rack at the handlebars when installing the rake

•

Check for proper seating and safe position of the glassware before lifting the rack from the

•

carrier