7 Maintenance and repairs

43

K-425 / K-436 Operation Manual, Version A

7 .2 .1

Breakage of glassware inside a housing chamber

Under rare conditions charged sample tubes or other glassware might break inside a housing
chamber. In such a case strictly follow the cleaning instructions below!

Cleaning under safe conditions

Switch off the SpeedDigester and unplug the power cord.

•

➡ Let the system cool down completely!

Carefully remove the installed rack(s).

•

Wear safety gloves to dispose of the defective glassware of the rack.

•

Use a long forceps to remove residual bits of glass off the housing chamber(s).

•

Use a long forceps to dab residual liquid and other components off the chamber(s).

•

Wear cut- and acid-safe gloves to wipe out the chamber(s) with wet damp cloth.

•

Check the surface below the instrument and clean if necessary.

•

Wait for the system to completely dry before reconnecting the device to mains!

7 .2 .2

Liquid ingress into the instrument

In case of liquid ingress into the housing the instrument is not safe to use anymore. Follow the instruc-
tions below to safely bring the instrument into operable condition again.

Switch off the SpeedDigester and unplug the power cord.

•

Use a long forceps to dab liquid off the outer housing parts. Wear safety gloves!

•

Carefully remove rack(s) if installed.

•

➡ Let the system cool down completely!

Use a long forceps to dab residual liquid off the chamber(s).

•

Wear cut- and acid-safe gloves to wipe out the chamber(s) with dry cloth.

•

Use a long forceps to clean the chamber(s) with a wet damp cloth.

•

➡ Call service! Do not reconnect the system to mains!

!

WARNING

Death or serious burns by electric current.

Switch off the instrument

•

Disconnect the power cord and prevent unintentional restart

•

All parts must be completely dried before the system can be checked by a service technician!
An electrical safety test and a functional test must be performed before the system can be used again.