MODULE UNCHARGED

During the stimulation a module may be uncharged. In this case the symbol for an uncharged battery
appears and an arrow flashing on the channel where the problem was detected (in this case, channel 1).
• Stop the stimulation and recharge the uncharged module.
• Abandon the uncharged module and continue the stimulation session without it.

THE MODULE WILL NOT PAIR WITH THE REMOTE CONTROL

At the time of first use, if the remote control is unable to pair all modules, an error message may appear.
• Ensure that the module is charged, and repeat the pairing step.
• If despite this the message returns, contact customer service provided and approved by Compex.

STIMULATION DOES NOT PRODUCE THE USUAL SENSATION
• Check that all settings are right and check electrodes are properly positioned.
• Change the positioning of the electrodes slightly.

STIMULATION CAUSES DISCOMFORT
• Electrodes lose their adhesive capacity and no longer provide adequate contact with skin.
• Electrodes are worn and must be replaced.
• Change the positioning of the electrodes slightly.