RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER CONCERNING

ELECTROMAGNETIC IMMUNITY

Compex Stimulator is designed for use in the electromagnetic environment stipulated below. The buyer or user of the

Compex Stimulator must ensure it is used in this recommended environment.

IMMUNITY TEST

TEST LEVEL IEC 60601

OBSERVANCE LEVEL

ELECTROMAGNETIC

ENVIRONMENT -

RECOMMENDATIONS

Electrostatic discharge (DES)

CEI 61000-4-2

±6 kV at the contact

±8 kV in air

±6 kV at the contact

±8 kV in air

Floors must be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material the relative
humidity must be maintained
at a minimum of 30%.

Fast transient electrical

bursts

CEI 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

±2 kV for power supply lines

Not Applicable for input/

output lines

The quality of the electrical
power supply should be that
of a typical commercial or
hospital environment.

Surge

CEI 61000-4-5

±1 kV differential mode

N/A

±1 kV differential mode

±2 kV joint mode

The quality of the power
supply should be that of
a typical commercial or
hospital environment.

Voltage dips, short

interruptions and voltage

variations on power supply

lines

CEI 61000-4-11

<5 % VT (dips >95 % de UT)

for 0.5 cycle

<40 % VT

(dips >60 % de UT)

for 5 cycles

<70 % VT

(dips >30 % de UT)

for 25 cycles

<5 % VT (dips >95 % de UT)

for 5 seconds

<5 % VT (dips >95 % de UT)

for 0.5 cycle

<40 % VT

(dips >60 % de UT)

for 5 cycles

<70 % VT

(dips >30 % de UT)

for 25 cycles

<5 % VT (dips >95 % de UT)

for 5 seconds

The quality of the power
supply should be that of
a typical commercial or
hospital environment. If the
Compex Stimulator user
requires continuous operation
during mains power cuts, it is
recommend that the Compex
Stimulator is powered by a
UPS or a battery.

Magnetic field at grid
frequency (50/60 Hz)

CEI 61000-4-8

3 A/m

3 A/m

Magnetic fields at the mains
frequency should be at a
level characteristic of a
typical location in a typical
commercial or hospital
environment.

NOTE :VT is the AC supply voltage before application of the test level.