8

Maintenance

DANGER!
Mains voltage
– Before beginning with maintenance work, unplug the

mains plug and protect the device against being unin-

tentionally switched back on again!

– Indicate maintenance work with warning signs.
– Maintenance work may only be performed by special-

ised personnel.

– Put safety equipment back into operation after mainte-

nance or repair work.

We recommend keeping a safety logbook Ä Chapter 13

‘Safety logbook’ on page 52, where all work (maintenance,

repairs......) performed on the device is entered.

The most important element of maintenance is regular

cleaning:

When cleaning the complete device, adhere to the

guidelines of the Accident Prevention Regulation (BGV

A3) ‐ especially if the device has been set up in a dusty

environment or if the processed source material pro‐

duces dust.

Functional part

Task

Test

Maintenance interval

Safety lock

Hood lock

Is the closed hood locked in

place when the main

switch is off?
If this test is failed, do not

continue to work until the

fault has been rectified.

Before each use

LID LOCK ACTIVATED light

on the control panel

Shows that the hood is

closed

When the hood is closed

the LID LOCK ACTIVATED

light should light up green

Once a month

Rotating bearings

Permanent lubrication

Bearing clearance

Every 2,000 h or annually

Drive motor

Permanent lubrication

Bearing clearance

Every 4,000 h or annually