background image

Functional part

Task

Test

Maintenance interval

V-belt

Motor to planetary disc

Check the voltage; remove

the housing; the belt must

not slacken by more than

approx. 10 mm when

pushed with your thumb.

Once a year

Fan

Grinding chamber cooling

and electronics

Proper function, clean

when soiled

Twice a year

Spindle clamping device (5)

for the grinding bowl

clamping device

Clamping the grinding

bowls

Mobility, oil if necessary

Every 1,000 h

Grinding bowl holder

Cork disc, rubber of the

pressure piece and grinding

bowl lid seal

Signs of use; replace if

pressed flat and no longer

elastic

Every 1,000 h