The plunger has 2 different sides for feeding the sample material in the

funnel to the grinding chamber. On the one hand, the smooth, round

side is suitable for finer material. On the other hand, the cross-shaped,

thinner side is suitable for long, fibrous material, like straw.

6.3 Comminution procedure with the protected funnel

1.

Switch the control switch (19) on the back of the device to AUTO.

2.

Close the cutting mill (see Ä Chapter 6.1.6 ‘Closing the cutting

mill’ on page 34).

3.

Push in the collecting vessel (8).

4.

Pull the sample pusher (24) out fully.

5.

Move the plunger (23) into the lower position.

6.

Flip open the funnel lid (21), add some comminution material and

close the lid again.
The quantity of comminution material is based on the particle feed

size and the grindability of the comminution material. It is best to

start with small quantities and increase them depending on the

success of the comminution.

7.

Switch on the deviceà press the green Start (34) button (see

Ä

Chapter 5.2 ‘Switching on’ on page 26).

8.

Move the sample pusher (24) down towards the plunger (23) until

the comminution material is in front of the plunger (23).

9.

Pull the plunger (23) upwards and move the sample pusher (24)

fully downwards until the comminution material falls into the cut-

ting chamber.

ð a slight operating noise becomes audible.

Using the device