Using the device – FRITSCH PULVERISETTE 19 User Manual

Page 40

Advertising
background image

2.

Attach one end of the connection hose (28) with the rubber sleeve

(29) to the adapter (27) and the other end to the cyclone sepa-

rator (30).

NOTICE!
There are 2 opposing hose connections on the

cyclone separator (30). Do not use the sealed con-

nection; leave it closed. This would increase the noise

level!

3.

Close the cutting mill with mounted standard funnel (1) (see

Ä

Chapter 6.1.6 ‘Closing the cutting mill’ on page 34).

4.

Screw the sample bottle (31) onto the adapter (33) on the bottom

of the cyclone separator (30); make sure the connection hose is

firmly connected and switch on the exhaust system.

5.

Switch on the cutting mill (see Ä Chapter 5.2 ‘Switching on’

on page 26).

6.

Add a little comminution material by hand and observe the oper-

ating noises. If the motor speed decreases audibly, reduce the

supply of comminution material. A fast vortex of comminution

material must form in the collection bottle.

Using the device

- 40 -