3 pre-crushing settings for coarse samples, 5 adjusting the vertical pressure scale, 1 adjusting – FRITSCH PULVERISETTE 2 User Manual

Page 25: Ä chapter 6.5.1 ‘adjusting, Using the device

Advertising
background image

6.4.3 Pre-crushing settings for coarse samples

To pre-crush coarse samples of up to 8 mm, the adjusting nut (c) can be

raised above the zero position if necessary so that coarse particles fit

below the pestle and can be comminuted. (see the illustration under

Ä

Chapter 6.4.1 ‘Positioning the pestle against the bottom of the mortar’

on page 24) For subsequent fine grinding, the pestle is lowered again to

the desired position. (Set to contact pressure [daN Downforce] between

2.5 - 17.5 daN)

6.5 Adjusting the vertical pressure scale

On new devices that are delivered with a mortar and pestle, the scale is

adjusted to fit the grinding parts and does not need to be readjusted

during commissioning.
The scale has to be readjusted if different or new grinding parts are used.

6.5.1 Adjusting

1.

Install the empty mortar (without grinding stock) and fasten it to

the carrier plate (see Ä Chapter 6.3.1 ‘Installing/removing the

mortar’ on page 22)

2.

Insert the pestle into the latch bearing in the lid (see

Ä

Chapter 6.3.2 ‘Installing/removing the pestle’ on page 23).

3.

Close the lid (1) with the latch (5)!

4.

Unscrew the adjusting nut (c) until the pestle is lifted from the

bottom of the mortar.

5.

Unscrew the adjusting screw (f) for the lateral pressure setting

until it stops.

6.

The pestle is moved to the grinding chamber centre.

Using the device

- 25 -

Advertising