Calibration, 11hi00 – A&D Weighing HC-3Ki User Manual

Page 29

Advertising
background image

27

8. CALIBRATION

‰

Calibration of the scale is required when it is initially installed, if it is moved often, or

it is moved a substantial distance. Calibration is also necessary in regular scale
maintenance due to normal mechanical wear-and-tear, changes in seasonal
temperature, humidity, air pressure, etc.

The scale is equipped with gravity compensation, which allows it to be calibrated in
one location and then adjusted to match the gravity acceleration at another location
where it will be used. But don’t worry about this, as far as you calibrate the scale
using a calibration weight and use it at same place.

The scale must perform "warm up" for at least 30 minutes before starting
calibration.


8-1. Calibration Procedure Using a Weight

The scale should be powered on at least one-half hour to warm it up before starting
the calibration procedure.

1. Remove the calibration switch cover, and press the calibration (CAL) switch.
The scale shows “Cal” in the count display.













‰

Press the

ON/OFF

key to exit

without calibrating the scale.


‰

Press and hold the

PRINT

key and

press the

ON/OFF

key, then you can

also enter calibration mode.




2. Press the

ZERO

key to enter into zero &

span calibration mode.


‰

The display flashes the required calibration

weight value.

Calibration (CAL)

switch

Bottom view

Calibration (CAL)

switch cover

888

Cal

02

COUNT

WEIGHT UNIT WEIGHT

88-

0.000

8

8

11Hi00

888

Cal

10

COUNT

WEIGHT UNIT WEIGHT

88

Cal

0

0

8

8

0.0

6.000

0

00

Advertising
This manual is related to the following products: