Charging, Press shock button – Defibtech DDU-2450 Series User Manual User Manual

Page 29

Advertising
background image

5.3 Selecting Energy

The user may select the desired energy level by pressing the softkey associated with
the Energy Select Icon (upper button). The default energy when entering Manual
Mode is always 150 J (adult pads) or 50 J (pediatric pads). Pressing the Energy Select
Icon will cycle the energy through the following options:

• 25, 50, 70, 100, 150, 200 Joules (Adult pads attached)
• 25, 50, 70, 100 Joules (Pediatric non-attenuating pads attached)
• 50 Joules (not selectable) (Pediatric attenuating pads attached)

5.4 Initiating Charge

To initiate charging, press the softkey associated with the CHARGE Softkey Icon
(middle button). The AED’s main screen will display a progress bar and produce an
audible rising tone while charging. The CHARGE Softkey Icon will change to the
DISARM Softkey Icon.

Charging

DISARM Softkey Icon

Charging Screen

While charging, pressing the softkey associated with the DISARM Softkey Icon
(middle button) will cancel the charge.

5.5 Delivering The Shock

When charging is complete, the AED will prompt the user to press the flashing SHOCK button and will produce
a two-toned audible alarm.

Manual Mode

Press SHOCK Button

The user should follow the AED’s onscreen text instructions and
press the flashing SHOCK button. (Note: If the SHOCK button is not
pressed within 30 seconds, the AED will automatically disarm.)

IMPORTANT: The DDU-2000 Series AED will not automatically
deliver a shock — the user must press the flashing SHOCK
button.

The user can abort charging or shock delivery at any time while
in Manual Mode by pressing the DISARM softkey (middle button)
or by pressing and holding down the AED’s ON/OFF button for
approximately two seconds to turn the unit off.

Shock Delivery Screen

SHOCK button (flashes when ready to shock)

5.
Manual Mode

(DDU-240

0 only)

29

DAC-E2510EN-BB

Advertising
This manual is related to the following products: