CPR Breathing

The CPR breathing option allows the user to enable or disable the CPR breathing coaching prompts
during CPR.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to CPR breathing:

AED Main Menu Ú Rescue options Ú CPR breathing

What it does: CPR breathing allows the user to enable/disable CPR breathing coaching prompts.

Use the TOP and BOTTOM softkey buttons to select the desired mode. When the desired selection is
chosen, press the CENTER softkey button to set the feature. The AED will now use that coaching setting.
Note: The factory default setting of CPR breathing is set to “disabled.”

Rescue Protocol

The AED supports two rescue protocols at one time. The Rescue protocol option allows the user to
select a rescue protocol. Rescue protocol options include the AHA/ERC 2010 protocol or “Custom.”
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to Rescue protocol:

AED Main Menu Ú Rescue options Ú Rescue protocol

What it does: The Rescue protocol option allows the user to select between up to two rescue protocols
that have been enabled in the unit. The factory default setting of the rescue protocol is AHA/ERC 2010.

To change the protocol, press the CENTER softkey button to highlight the protocol. The user will be
prompted to enter a password to proceed. The password may be obtained from your medical director
or from Defibtech. (For Defibtech contact information, please refer to the

Contacts

section in

Chapter 14.) Once the password has been entered, the user may select between the two protocols.

To enter the password use the TOP softkey button to scroll through numbers. Once the correct number
appears, use the CENTER softkey button to move to the next space. Once all of the numbers have been
entered press the CENTER softkey button. The user will now be able to choose a different rescue protocol.

Settings

Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to Settings:

AED Main Menu Ú Rescue options Ú Settings

What it does: The Settings option allows the user to change the currently enabled
protocol by entering a special protocol code. This code is a special code that encodes
all the important information regarding the protocol. This code is custom generated by
Defibtech. If the code is not entered correctly the protocol will not be changed. Based on
the protocol code entered, the currently selected protocol will be changed to that described
by the special protocol code. You can obtain this code from your medical director or from
Defibtech. (For Defibtech contact information, please refer to the

Contacts

” section in

Chapter 14.) Once the code has been entered, the settings will have been changed.

To enter the code use the TOP softkey button to scroll through numbers/letters. Once
the correct number/letter appears, use the CENTER softkey button to move to the next
space. Once all of the numbers/letters have been entered, press the CENTER softkey
button. The settings will have been changed based on the code entered.