7.6 Troubleshooting

The following table lists the symptoms, the possible causes, and the possible corrective actions for common problems.
Refer to the other sections of the user manual for detailed explanations on how to implement the corrective actions.
If the unit continues to be non-functional, refer the unit for servicing. (Refer to

Chapter 14

of this manual for contact

information.)

Symptom

Possible Cause

Corrective Action

Unit will not turn on

Battery pack not inserted

Insert battery pack

Battery pack depleted or
needs servicing

Replace battery pack
or call for service

Unit needs servicing

Call for service

Unit immediately turns off

Battery pack depleted

Replace battery pack

Unit needs servicing

Call for service

ASI flashes red and/or unit makes periodic
”beep” sound

Unit needs servicing

Go to AED Status
Screen by pressing the
CENTER softkey button
or call for service

Battery pack non-functional

Replace battery pack

Defibrillation pads are not
pre-connected to unit

Connect defibrillation
pads to unit

Defibrillation pads
or battery pack expired

Replace expired
component

ASI does not flash at all

Battery pack not inserted

Insert battery pack

Battery pack is low
or needs servicing

Replace battery pack or
call for service

Unit needs servicing

Call for service

“Power on test failed, service code ‘xxxx’”
prompt

Unit needs servicing

Record code number
and call for service

“Battery test failed, service code ‘xxxx’”
prompt

Battery pack needs servicing

Record code number
and call for service

“Service required” prompt

Unit needs servicing

Call for service

“Replace battery now” prompt

Battery pack capacity is
critically low

Unit may not deliver a
shock, replace battery
pack immediately

“Battery low” prompt

Battery pack capacity is
getting low

Replace battery pack as
soon as possible

Display screen does not work

Battery pack depleted

Replace battery pack

Battery pack not inserted
properly

Make sure battery pack
is oriented correctly and
fully inserted

Unit needs servicing

Call for service

“ Plug in pads connector” prompt

Pads connector not plugged in

Plug in pads connector

Pads connector broken

Replace pads

Unit’s connector broken

Call for service

7.

Maintenance

And

Tr

oubleshooting

38