10 Event Viewing

This chapter includes information about DefibView, Defibtech Data Cards (DDC cards), and downloading internal data logs.

10.1 DefibView

DefibView is a Windows-based software application that reads data stored on a DDC card or downloaded through
the USB port and displays the data on a personal computer. DefibView serves the following primary functions:

• Enables emergency care personnel to review a cardiac episode from the time the AED was turned on and

connected to the patient until the unit is turned off.

• Provides maintenance personnel with additional parameter information to assist in troubleshooting a device

suspected of malfunctioning.

DefibView is a stand-alone software application. DefibView cannot be used while the AED is in operation, and its
function is solely to support post-event review.

CAUTION

Not intended for clinical use. Information presented by
DefibView should not be used for making clinical decisions.

10.2 Defibtech Data Cards (DDC Cards)

If a DDC card is installed in the unit, every time the DDU-2000 Series AED is turned on, the following information is
recorded on a new file on the card:

• The time the AED was turned on.
• Other data such as: ECG data, time data, audio data (audio enabled cards only), event milestones such as:

motion detection, shock advice, shock delivery information.

This information can be reviewed using the DefibView application.

CAUTION

Using non-Defibtech Data Cards
(DDC cards) may damage the unit
and will void the warranty.

10.3 Downloading The Internal Data Log

Regardless of whether a DDC card is installed in the unit, select information is recorded internally within the
DDU-2000 Series AED. The information recorded is limited to:

• The time the AED was turned on.
• Other data such as event milestones (motion detection, shock advice, shock delivery information, etc.).
• Important ECG information.
Note: Audio data is not logged internally.

Downloading The Internal Data Log Using The DDC Card

To download the internally logged information, perform the following procedure:

• Insert a DDC card into the unit.
• Turn the unit on in Maintenance Mode by pressing the center softkey button.
• Press the tool icon to enter the AED Maintenance screen.
• From the AED Maintenance screen select the Transfer Data To Card option.
• Allow the unit to write the contents of the internal log to the DDC card.

The DDU-2000 Series AED will write the contents of the internal log onto the DDC card. This information can then
be reviewed using DefibView software.

Downloading The Internal Data Log Using The USB Port

To download the internal data log using the USB port in the unit, connect a USB cable between the unit and a PC.
Launch the DefibView software and follow USB download directions.