Shockable Rhythm Criteria

When placed on a patient meeting the indications for use criteria, the DDU-2000 Series AED is designed to recommend
a defibrillation shock when it detects proper pad impedance and one of the following:

Ventricular Fibrillation: Peak-to-peak amplitude at least 200 µVolts.

Some very low amplitude or low frequency VF rhythms may not be interpreted as shockable.

Ventricular Tachycardia (Including ventricular flutter and polymorphic VT):
Cardiac rhythm rate of at least 180 bpm and peak-to-peak amplitude at least 200 µVolts.

Some very low amplitude or low frequency VT rhythms may not be interpreted as shockable.

The DDU-2000 Series AED is designed to recommend no shock for all other rhythms, including Normal Sinus
Rhythms, fine Ventricular Fibrillation (<200 µVolts), and some slow Ventricular Tachycardia and Asystole.

Patient Analysis System Performance

Rhythm Class

ECG Test
Sample

1

Size

Algorithm Performance

1

Specifications

Performance

2

90% Lower
Confidence Limit

2

Shockable Rhythm –
Ventricular Fibrillation

227

>97%

>95%

Meets the AAMI
DF80 requirement
and AHA
recommendation

2

of

Sensitivity > 90%

Shockable Rhythm –
Ventricular Tachycardia

101

99%

>97%

Meets the AAMI
DF80 requirement
and AHA
recommendation

2

of

Sensitivity > 75%

Non-Shockable
Rhythm – Normal
Sinus Rhythm

213

100%

100%

Meets the AAMI
DF80 requirement
of Specificity
>95% and the AHA
recommendation

2

of

Specificity > 99%

Non-Shockable
Rhythm – Asystole

113

100%

100%

Meets the AAMI
DF80 requirement
and the AHA
recommendation

2

of

Specificity > 95%

Non-Shockable
Rhythm – All other
non-shockable
rhythms

248

>99%

>98%

Meets the AAMI
DF80 requirement
and the AHA
recommendation

2

of

Specificity > 95%

1. From Defibtech ECG Rhythm Databases.

2. Automatic External Defibrillators for Public Access Defibrillation: Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety.

American Heart Association (AHA) Task Force on Automatic External Defibrillation, Subcommittee on AED Safety and Efficacy. Circulation, 1997;95:1677-1682.

Note: Additional information available upon request.

WARNING

WARNING

11

.

Tec

hnical Specifications

Defibtech DDU-2000 Series AED (continued)

57