Connector for standard load cells, Connector for digital load cells and digital i/o – Flintec II IPC User Manual
Page 9

FlintWeigh II IPC Technical Manual, Rev. 1.06 November 2011
Page 9 of 36
4.1.1.
Connector for Standard Load Cells
The connectors J1 and J2 can be used for connecting standard load cells in 4– or 6–wire technique.
Hint: If 4–wire technique will be used, then pin no. 1 and no. 2 as well as pin no. 4 and no. 5 have to be short-
circuited.
It is mandatory to interconnect the shield of the connection cable with one of the shield connector pins
(pin no. 3, 6 or 9).
Pin no.
Signal
Description
1
+ Excitation
Excitation, positive pol
2
+ Sense
Sense, positive pol
3
Shield
Shield connector for load cell connection cable
4
– Sense
Sense, negative pol
5
– Excitation
Excitation, negative pol
6
Shield
Shield connector for load cell connection cable
7
+ Signal
Measurement signal, positive pol
8
– Signal
Measurement signal, negative pol
9
Shield
Shield connector for load cell connection cable
Tab. 4.1: Plug connector for standard load cells (J1 / J2)
4.1.2.
Connector for Digital Load Cells and Digital I/O
The connector J3 provides a RS485 bus interface and a power supply for the digital load cells.
IMPORTANT: It is mandatory to leave pin no. 2 unconnected, if the digital load cells are powered externally,
see also chapter 4.3.
The connector also provides the digital I/O interfaces,see chapter 4.4.
Pin no.
Signal
Description
1
GND
Power supply
+12V DC / 450 mA
2
+12V DC
3
RS485 A
RS485 bus – non-inverted Signal
4
RS485 B
RS485 bus – inverted signal
5
IO - GND
Digital I/O – signal ground
6
IO - In1
Digital I/O – input
7
IO - Out1
Digital I/O – output 1
8
IO - Out2
Digital I/O – output 2
Tab. 4.2: Plug connector for digital load cells and digital I/O (J3)
Fig. 4.3: Connector pin assignment J3