KERN & SOHN ABJ 320-4NM User Manual

Page 79

Advertising
background image

ABS-N_ABJ-NM-BA-e-1312

79

4. Connect filter

Press the navigation buttons ( ) repeatedly

until „NO.FIL“ is displayed. The current setting is marked

by the stability display ( ).

with ( )

Weighing value is filtered

without ( ) Weighing value is not filtered

Change using TARE

5. Return to weighing mode

Press ON/OFF repeatedly or for 3 s.

With activated menu setting „MANU“ (see step 3, with )
the indicator

will appear.

With deactivated menu setting „MANU“ (see step 3,

without ) starts the continuous data output

automatically, step 7 is not applied.

6. Tare weighing container

If necessary, place an empty container on the balance and tare it by pressing

TARE

.

Place weighed goods and wait until the stability display ( ) appears. The

weighing value is issued automatically.

7.

Press PRINT (only with menu setting „MANU ON

)

The continuous data output starts, the indicator

goes out.

8. Place goods to be weighed on balance

Every display change is continuously issued,

(data issue interval approx. 100 msec).

Interrupt continuous data output
ON-OFF

= interrupt

PRINT

= restart

Advertising
This manual is related to the following products: