Ultrasonic milk analyser

Operation manual

23.08.10

35/98

in the milk analyser as Calibration 1 - cow) and press the button Enter.The
following message appears on the display:

Calibration 1

Getting ready


6.4.2.2. Entering values for the separate sample parameters
Wait for the sound signal and the following menu on the display:

Cal1 Samp High
Fat=f.ff
SNF=y.yy
Den=d.dd
------------------------------
Lac=l.ll
Sol=s.ss
Pro=p.pp
Exit


In this display is possible to be entered the results from the milk with high
FAT
chemical analyses, which were received by the methods from Appendix
Methods.
For example:

- by using the buttons “up” ▲ and ”down” ▼, choose the needed

parameter to be entered;

- by pressing “Enter” the cursor will mark the number to be changed. For

example: FAT=01.29;

- using the buttons “up” ▲and ”down” ▼, enter the needed value;
- by next pressing of Enter pass towards the next number;
- after completion of entering the needed FAT value, press Enter;
- the cursor returns to its starting position.

Cal 1 Samp.High
Fat=f.ff
SNF=y.yy
Den=d.dd